咨询与建议

看过本文的还看了

相关文献

该作者的其他文献

文献详情 >Consistency test of PMT SPE sp... 收藏

Consistency test of PMT SPE spectrum from dark-noise pulses and LED low-intensity light

作     者:Yu-ting Wei Meng-yun Guan Wei-xing Xiong You-yu Gan Chang-gen Yang 

作者机构:Key Laboratory of Particle AstrophysicsInstitute of High Energy PhysicsChinese Academy of ScienceBeijingChina School of PhysicsUniversity of Chinese Academy of ScienceBeijingChina 

出 版 物:《Radiation Detection Technology and Methods》 (辐射探测技术与方法(英文))

年 卷 期:2018年第2卷第1期

页      面:76-79页

学科分类:080901[工学-物理电子学] 0809[工学-电子科学与技术(可授工学、理学学位)] 08[工学] 080401[工学-精密仪器及机械] 0804[工学-仪器科学与技术] 0803[工学-光学工程] 

基  金:The study was supported by China Ministry of Science and Technology under Grant No.2016YFA0400304 

主  题:PMT SPE Dark noise LED Waveform 

摘      要:Photomultiplier tube(PMT)has been an essential instrument for low background experiments,and the accuracy of single photoelectron(SPE)calibration will directly affect the energy reconstruction of the *** SPE spectrum can be acquired from the charge distribution of dark-noise pulses or the low-intensity light signal triggered by ***,there are arguments on whether the two methods are *** this paper,the Hamamatsu R5912-20mod PMT and the EMID642KNFL PMT have been tested adopting the two methods with signal waveform recording *** comparison,traditional charge-to-digital conversion system is also used to test SPE *** results show that they are consistent when the PMTs are working at typical recommended high voltage(HV).

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分