Abnormal Blood Glucose Concentration on Degradation Behavior of AZ31 Magnesium Alloy
AZ31 镁合金的降级行为上的反常的血葡萄糖集中作者机构:School of Chemistry and Chemical EngineeringShandong UniversityJinan 250100P.R.China Institute of Marine Science and TechnologyShandong UniversityQingdao 266237 P.R.China
出 版 物:《Chemical Research in Chinese Universities》 (高等学校化学研究(英文版))
年 卷 期:2020年第36卷第6期
页 面:1217-1226页
核心收录:
学科分类:08[工学] 080502[工学-材料学] 0805[工学-材料科学与工程(可授工学、理学学位)]
基 金:Supported by the Self Innovation Project of Shandong Province China(No.012CX80106)
主 题:Medical implant material AZ31 magnesium alloy Degradation Blood glucose Hyperglycemia
摘 要:In order to evaluate the effect of blood glucose concentration on the reliability of AZ31 magnesium alloy medical implant,the corrosion of AZ31 alloy was studied in a simulated physiological saline *** is found that when the glucose concentration is in the normal range of ca.1g/L,the corrosion of AZ31 alloy can be inhibited However,when the glucose concentration becomes higher like that of diabetic patients,the degradation of AZ31 alloy is significantly ***,for diabetic patients,the change in glucose concentration must be taken into consideration in order to ensure the reliability ofAZ31 medical implants.