The Dosimetric Characteristics and Potential Limitation in Clinical Application of a Low Energy Photon Intra-Operative Radiotherapy System

The Dosimetric Characteristics and Potential Limitation in Clinical Application of a Low Energy Photon Intra-Operative Radiotherapy System

作     者:Zhenhua Xiao Ouyang Bin Zhenyu Wang Botian Huang Bixiu Wen 

作者机构:Department of Radiation Oncology The First Affiliated Hospital of Sun Yat-sen University Guangzhou China 

出 版 物:《International Journal of Medical Physics, Clinical Engineering and Radiation Oncology》 (医学物理学、临床工程、放射肿瘤学(英文))

年 卷 期:2015年第4卷第2期

页      面:184-195页

学科分类:1002[医学-临床医学] 100214[医学-肿瘤学] 10[医学] 

主  题:Intra-Operative Radiotherapy Low Energy Photon Dosimetry Isotropy Limitation 

摘      要:Purpose: To investigate the dosimetric characteristics of a low energy photon intra-operative radiotherapy (IORT) system and explore its potential limitation in clinical application. Methods: A special water phantom, a parallel-plate ionization chamber and an electrometer were used to measure the depth dose rate, isotropy of dose distribution in X/Y plane, dosimetry reproducibility of bare probe and spherical applicators of different size which were used in comparison with the system data. Results: The difference in depth dose rate between the measurement and system data for bare probe is -2.16% ± 1.36%, the range of the relative deviation for isotropy in the X/Y plane is between -1.9% and 2.1%. The difference in depth dose rate, transfer coefficient, isotropy in X/Y plane between the measurement and system data for the whole set of spherical applicators is -10.0% - 2.3%, -8.9% - 4.2% and -1.6% - 2.6%, respectively. Higher surface dose rate and steeper gradient depth dose are observed in smaller spherical applicators. The depth dose rate and isotropy for bare probe and spherical applicators have been shown good reproducibility. The uncertainty of measurement is associated with the positioning accuracy, energy response, noise current and correction function f’(R). Conclusions: Thorough commissioning of the low energy photon IORT system helps us better understand the dosimetry characteristics, verify the system data, obtain adequate data for clinical application and routine quality assurance. The steep gradient depth dose and limited treatment range may restrain its potential in clinical application.

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分