Preparation Process and Mechanical Properties of PLA/PCL Intravascular Stent

Preparation Process and Mechanical Properties of PLA / PCL Intravascular Stent

作     者:WANG Cong-er 

作者机构:Key Laboratory of Textile Science & Technology Ministry of Education Donghua University Shanghai 201620 China College of Textiles Donghua University Shanghai 201620 China 

出 版 物:《Journal of Donghua University(English Edition)》 (东华大学学报(英文版))

年 卷 期:2014年第31卷第5期

页      面:640-642页

核心收录:

学科分类:08[工学] 080502[工学-材料学] 0805[工学-材料科学与工程(可授工学、理学学位)] 

基  金:"111 Project " Biomedical Textile Materials Science and Technology China 

主  题:biodegradable intravascular stents polylactic acid(PLA) heat setting process compression properties 

摘      要:A hand-made biodegradable intravascular stent with zig- zag structure has been developed, which made from polylactic acid ( PLA ) braided thread and poly-eaprolactone ( PCL ) tube. This paper showed the preparation process of the PLA braded thread and stent, and discussed the heat setting process of the stents. The sixteen groups of stents were made for the heat setting experiments, discussing two factors on four levels. Radial compression behavior of stents was tested by plate compression method. The result showed that both the heat setting temperature and time had significant effects on stent compression strength. The compression recovery rate was mainly affected by heat setting temperature. The optimized heat setting condition should be 90-100℃ for 45 rain.