咨询与建议

看过本文的还看了

相关文献

该作者的其他文献

文献详情 >Amperometric Determination of ... 收藏

Amperometric Determination of Serum Cholesterol with Pencil Graphite Rod

Amperometric Determination of Serum Cholesterol with Pencil Graphite Rod

作     者:Nidhi Chauhan Jagriti Narang Chandra S. Pundir 

作者机构:不详 

出 版 物:《American Journal of Analytical Chemistry》 (美国分析化学(英文))

年 卷 期:2010年第1卷第2期

页      面:41-46页

学科分类:1002[医学-临床医学] 100214[医学-肿瘤学] 10[医学] 

主  题:Cholesterol Cholesterol Oxidase Pencil Graphite Enzyme Electrode Serum Amperometric Biosensor 

摘      要:A cholesterol oxidase from Streptomycin sp. was immobilized onto pencil graphite rod and employed for amperometric determination of serum cholesterol. The method has the advantage over earlier amperometric methods that it requires low potential to generate electrons from H2O2, which does not allow ionization of serum substances. The optimum working conditions of amperometric determination were pH 6.8, 25?C and 30 s. The current measured was in proportion to cholesterol concentration ranging from 1.29×10-3 to 10.33×10-3 M. Minimum detection limit of the method was 0.09 ×10-3 M. Mean analytical recovery of added cholesterol (100 mg/dl and 200 mg/dl) in serum was 85.0% & 90.0% respectively. Within batch and between batch coefficients of variations were 1.59% & 4.15% respectively. A good correlation (r = 0.99) was obtained between serum cholesterol values by standard enzymic colorimetric method and the present method. No interference by metabolites was observed in the method. The enzyme electrode was reused 200 times over a period of 25 days, when stored at 4?

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分