The Clinical Randomized Controlled Trial on the Safety and Efficacy of Polyethersulfone Membrane Dialyzer in the Maintenance of Hemodialysis Patients

The Clinical Randomized Controlled Trial on the Safety and Efficacy of Polyethersulfone Membrane Dialyzer in the Maintenance of Hemodialysis Patients

作     者:GU Ying-li LIN Fu-jun ZHU Chun JIANG Geng-ru 

作者机构:Xinhua Hospital Affiliated to Shanghai Jiao Tong University School of Medicine Shanghai 200092 China 

出 版 物:《Chinese Journal of Biomedical Engineering(English Edition)》 (中国生物医学工程学报(英文版))

年 卷 期:2010年第19卷第4期

页      面:148-155页

学科分类:08[工学] 080501[工学-材料物理与化学] 0805[工学-材料科学与工程(可授工学、理学学位)] 080502[工学-材料学] 

主  题:hemodialysis polyethersulfone membrane dialyzer randomizedcontrolled trial solute blood phosphate 

摘      要:Objective: To observe the safety and efficacy of polyethersulfone membrane dialyzer used by chronic renal failure patients in clinical maintenance hemodialysis (MHD). Methods: From January to February 2009, 36 patients were recruited from Xinhua Hospital Affiliated to Shanghai Jiao Tong University School of Medicine. The experiment was approved by the ethnical committee of Xinhua Hospital and under permissions fronl patients. All the patients were randomized into polyethersulfone group (PES, n=18) and polysulfone group (PS, n=18). Hemodialysis was given by using Fresenius 4008S capacity-controlled dialyser and ultra-purified bicarbonate dialysate, more than 3.5 h each time, three times per-week, followed by heparin anticoagulation for a week, three times totally. Changes in blood urea nitrogen (BUN), serum creatinine (Set), phosphate (p2+), hemoglobin and albumin levels were determined for the efficacy and safety evahlation. Results: The serum BUN, Set and p2+ concentrations significantly reduced after hemodialysis with different dialyzers, and the decrease amplitude of two groups was equivalent(P〉0.05). The clearance rate of the serum BUN and Scr had no significant difference among two groups(P〉O.05). The clearance rate (ml/min) of the sermn P^2+ was 144.57 ±27.83 v.s. 117.15 ±22.77 in two groups. The clearance rate of the serum p^2+ in PES group was more than that in PS group, and the difference was significant between two groups (P=0.0001). It was indicated that PES and PS membrane could efficiently eliminate serum mieromolecule solute in MHD patients, so PES membrane excelled PS membrane in eliminating serum p2+. The hemodialysis safety index, for example serum hemoglobin, albumin and blood pressure, had no significant difference between the two groups before and after hemodialysis (P〉0.05). Conclusion: The efficacy and safety of the PES hollow fiber membrane dialyzer is equivalent to that of the imported PS membrane dialyzer in hemodialysis for MHD pat

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分