This study examined the image quality on Computed Radiography (CR) in conjunction with the proper radiation dose as indicated based on the set-up on the anode-cathode voltage (V) and filament current (i). During this ...
详细信息
This study examined the image quality on Computed Radiography (CR) in conjunction with the proper radiation dose as indicated based on the set-up on the anode-cathode voltage (V) and filament current (i). During this study, a set of radiographic images of step wedge for various combinations of V and i was developed. The results showed that good image quality may be achieved by considering the effective V and i set up. Increasing V had a better effect to the image quality rather than increasing i. The X-ray power set-up of 2.5-4.0 kW with voltage 50 kV-80 kV and current of 50 mA may be chosen depends on the targeted aluminum thickness equivalent to obtain good image quality at the acceptable low radiation dose. Optimization set-up based on aluminum (Al) thickness equivalent may lead to significant reduction on the radiation dose and power consumption.