Preliminary study on modelling, fabrication by photo-chemical etching and in vivo testing of biodegradable magnesium AZ31 stents
作者机构:Department of Mechanical and Materials EngineeringUniversity of CincinnatiOH45221USA Department of Chemical and Environmental EngineeringUniversity of CincinnatiOH45221USA Charles River Laboratories Montreal ULCBoisbriandQuebecJ7H 1N8Canada Medical Products Market ConsultingIncIndianapolisIN46202USA Waygate TechnologiesBaker HughesCincinnatiOH45241USA
出 版 物:《Bioactive Materials》 (生物活性材料(英文))
年 卷 期:2021年第6卷第6期
页 面:1663-1675页
核心收录:
学科分类:0831[工学-生物医学工程(可授工学、理学、医学学位)] 0710[理学-生物学] 08[工学] 0805[工学-材料科学与工程(可授工学、理学学位)] 0836[工学-生物工程]
基 金:The authors would like to acknowledge the financial support provided by the NSF ERC for Revolutionizing Biomaterials through grant EEC-EEC-0812348 We would like to thank Dr.Sarah Pixley for helping us with the sample fixation for the micro-CT analysis and for proofreading the manuscript The help of Dr.Melodie Fickenscher with the SEM imaging and EDS analysis of the corroded stents is appreciated
主 题:Magnesium alloy FEA modelling Stents Photo-chemically etching Micro-CT Histology
摘 要:Magnesium metal(Mg)is a promising material for stent applications due to its biocompatibility and ability to be resorbed by the *** of stents by laser cutting has become an industry *** alternative approach uses photo-chemical etching to transfer a pattern of the stent onto a Mg *** this study,we present three stages of creating and validating a stent prototype,which includes design and simulation using finite element analysis(FEA),followed by fabrication based on AZ31 alloy and,finally,in vivo testing in peripheral arteries of domestic *** to the preliminary character of this study,only six stents were implanted in two domestic farm pigs weighing 25-28 kg and they were evaluated after 28 days,with an interim follow-up on day *** left and right superficial femoral,the left iliac,and the right renal artery were selected for this *** diameters of the stented artery segments were evaluated at the time of implantation,on day 14 and then,finally,on day 28,by quantitative vessel analysis(QVA)using fluoroscopic *** Coherence Tomography(OCT)imaging displayed some malposition,breaks,stacking,and protrusion into the lumen at the proximal,distal,and mid-sections of the stented *** stents degraded with time,but simultaneously became embedded in the *** 28 days,the animals were euthanized,and explanted vessels were fixed for micro-CT imaging and histology ***-CT imaging revealed stent morphological and volumetric changes due to the in-body *** in vivo corrosion rate of 0.75 mm/year was obtained by the CT *** histology suggested no-life threatening effects,although moderate injury,inflammation,and endothelialization scores were observed.