Fedora-type magnetic compression anastomosis device for intestinal anastomosis

为肠的吻合的浅顶软呢帽类型磁性的压缩吻合设备

作     者:Huan Chen Tao Ma Yue Wang Hao-Yang Zhu Zhe Feng Rong-Qian Wu Yi Lv Ding-Hui Dong 

作者机构:National Local Joint Engineering Research Center for Precision Surgery&Regenerative MedicineThe First Affiliated Hospital of Xi'an Jiaotong UniversityXi’an 710061Shaanxi ProvinceChina Department of Hepatobiliary SurgeryThe First Affiliated Hospital of Xi’an Jiaotong UniversityXi’an 710061Shaanxi ProvinceChina 

出 版 物:《World Journal of Gastroenterology》 (世界胃肠病学杂志(英文版))

年 卷 期:2020年第26卷第42期

页      面:6614-6625页

核心收录:

学科分类:1002[医学-临床医学] 100210[医学-外科学(含:普外、骨外、泌尿外、胸心外、神外、整形、烧伤、野战外)] 10[医学] 

基  金:National Natural Science Foundation of China  NSFC  (81470896) 

主  题:Magnetic compression anastomosis Anastomotic stenosis Size of anastomat Compression pressure Fedora-type magnetic compression anastomosis device 

摘      要:BACKGROUND Although previous studies have confirmed the feasibility of magnetic compression anastomosis(MCA),there is still a risk of long-term anastomotic *** traditional MCA devices,a large device is associated with great pressure,and eventually increased *** To develop a novel MCA device to simultaneously meet the requirements of pressure and *** Traditional nummular MCA devices of all possible sizes were used to conduct ileac anastomosis in *** mean(±SD)circumference of the ileum was 13.34±0.12 *** on short-and long-term follow-up results,we determined the appropriate pressure range and minimum ***,we introduced a novel“fedora-typeMCA device,which entailed the use of a nummular magnet with a larger sheet *** With traditional MCA devices,the anastomoses experienced stenosis and even closure during the long-term follow-up when the anastomat was smaller thanΦ5 ***,the risk of leakage increased when it was larger thanΦ4 *** comparison of the different designs,it was found that the“fedora-typeMCA device should be composed of aΦ4-mm nummular magnet with aΦ6-mm sheet *** The diameter of the MCA device should be greater than 120%of the enteric *** novel“fedora-typeMCA device controls the pressure and optimizes the size.