咨询与建议

看过本文的还看了

相关文献

该作者的其他文献

文献详情 >Energy response improvement fo... 收藏

Energy response improvement for photon dosimetry using pulse analysis

Energy response improvement for photon dosimetry using pulse analysis

作     者:H.Zaki Dizaji 

作者机构:Faculty of ScienceImam Hossein Comprehensive University 

出 版 物:《Chinese Physics C》 (中国物理C(英文版))

年 卷 期:2016年第40卷第2期

页      面:74-78页

核心收录:

学科分类:0709[理学-地质学] 07[理学] 0708[理学-地球物理学] 0804[工学-仪器科学与技术] 0827[工学-核科学与技术] 0703[理学-化学] 0704[理学-天文学] 070202[理学-粒子物理与原子核物理] 0702[理学-物理学] 

主  题:personal dose equivalent photon dosimetry diode detector energy compensation 

摘      要:During the last few years,active personal dosimeters have been developed and have replaced passive personal dosimeters in some external monitoring systems,frequently using silicon diode *** photons interact with the constituents of the diode detector and produce *** photon-induced electrons deposit energy in the detector's sensitive region and contribute to the response of diode *** achieve an appropriate photon dosimetry response,the detectors are usually covered by a metallic layer with an optimum *** metallic cover acts as an energy compensating *** this paper,a software process is performed for energy *** data sampling based on pulse height is used to determine the photon dose *** method is applied to improve the energy response in photon *** detector design is optimized for the response function and determination of the photon dose *** personal dose equivalent is determined in the energy range of 0.3-6 *** error values of the calculated data for this wide energy range and measured data for ^133Ba,^137Cs,^60Co and ^241Am-Be sources respectively are up to 20%and 15%.Fairly good agreement is seen between simulation and dose values obtained from our process and specifications from several photon sources.

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分