Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake
作者机构:Department of RadiologyTongji HospitalTongji Medical CollegeHuazhong University of Science and TechnologyWuhan 430030China Department of Health RecordsTongji HospitalTongji Medical CollegeHuazhong University of Science and TechnologyWuhan 430030China
出 版 物:《Current Medical Science》 (当代医学科学(英文))
年 卷 期:2020年第40卷第1期
页 面:178-183页
核心收录:
学科分类:08[工学] 0812[工学-计算机科学与技术(可授工学、理学学位)]
主 题:tube voltage contrast agent coronary CT angiography radiation dose iodine intake
摘 要:The effect of low voltage and low concentration contrast agent on image quality of coronary CT angiography,radiation dose and iodine intake was evaluated.A total of 121 patients with body mass index(BMI)0.05).The radiation doses in groups A,B,C and D were 2.6±1.4,3.6±1.&4.9±3.5 and 4.9±2.8 mSv,*** radiation dose in group A was significantly less than that in the rest three groups(P0.05).The iodine intakes in groups A,B,C and D were 14.9±1.5,15.0±1.5,17.7±2.0