CAPILLARY ELECTROPHORESIS DETERMINATION OF INOSINE IN SOMEDRUG FORMULAE USING HYPOXANTHINE AS INTERNAL STANDARD
CAPILLARY ELECTROPHORESIS DETERMINATION OF INOSINE IN SOME DRUG FORMULAE USING HYPOXANTHINE AS INTERNAL STANDARD作者机构:Acad Sinica USTC Grad Sch Beijing 100039 Peoples R China
出 版 物:《Chinese Chemical Letters》 (中国化学快报(英文版))
年 卷 期:1995年第6卷第8期
页 面:701-704页
核心收录:
摘 要:A new rapid and accurate method for determining inosine by capillary zone electrophoresis(CZE) is presented. The electrophoresis separation was performed at 24℃ in a fused-silica capillary(51.2cm×50 μmi.d., 38.5cm to UV detector)using 20mmol/L borax-20mmol/L H3PO4(pH 8.5) as the background electrolyte at 24KV applied voltage. Inosine samples were led into the capillary electrokinetically at 10KV for 15 seconds. In the range of 20μg/mL to 300μg/mL, inosine can be quantitatively determined with a deduction limit of 2.1μg/mL and a RSD of 2.92%. Hypoxanthine was used as the internal standard.