REDUCTION IN PERITONITIS INCIDENCE IN CONTINUOUS AMBULATORY PERITONEAL DIALYSIS WITH A NEW CATHETER AND IMPLANTATION TECHNIQUE
REDUCTION IN PERITONITIS INCIDENCE IN CONTINUOUS AMBULATORY PERITONEAL DIALYSIS WITH A NEW CATHETER AND IMPLANTATION TECHNIQUE作者机构:Austin Diagnostic ClinicAustinTexas Bioengineering Program and the Department of Chemical EngineeringThe University of TexasAustinTexas University of Alberta Department of Medicine university of Calgary Department of Biological Sciences Moncief-Popovich Research InstituteAustinTexas USA
出 版 物:《Chinese Journal of Biomedical Engineering(English Edition)》 (中国生物医学工程学报(英文版))
年 卷 期:1993年第2卷第2期
页 面:86-86页
学科分类:0831[工学-生物医学工程(可授工学、理学、医学学位)] 08[工学] 0836[工学-生物工程]
主 题:peritonitis tunnel maintain circumstance failed biofilm cultured dialysis CAPD episode
摘 要:Recurrent episodes of peritonitis with the same organism cultured at thecatheter exit site suggest that early tunnel colonization is associated with a transferof these organisms through the catheter tunnel directly from the skin into the peri-toneal *** this circumstance the catheter has failed in its fundamental designto establish and maintain a bacteriological barrier.A three part study was conductedusing the Moncrief-Popovich Catheter and Implantation *** first part