Novel device for monitoring respiratory rate during endoscopy-A thermodynamic sensor
Novel device for monitoring respiratory rate during endoscopy-A thermodynamic sensor作者机构:Department of Gastroenterology and HepatologyCleveland Clinic FloridaWestonFL 33324United States Department of Internal MedicineCleveland Clinic FloridaWestonFL 33331United States Department of TechnologyArtep IncorporatedElliot CityMD 21042United States Department of Quantitative Health SciencesCleveland ClinicClevelandOH 44195United States Department of Gastroenterology and HepatologyDigestive Disease and Surgery InstituteCleveland ClinicClevelandOH 44195United States
出 版 物:《World Journal of Gastrointestinal Pharmacology and Therapeutics》 (世界胃肠药理与治疗学杂志(英文版)(电子版))
年 卷 期:2019年第10卷第3期
页 面:57-66页
学科分类:10[医学]
主 题:Sedation Capnography Linshom Thermodynamic sensor Endoscopy
摘 要:BACKGROUND Monitoring ventilation accurately is an indispensable aspect of patient care in procedural *** current gold standard method of monitoring ventilation is by measuring exhaled carbon dioxide concentration,known as capnography.A new device utilizing thermodynamic measurement,the Linshom Respiratory Monitoring Device(LRMD),has been designed to measure respiratory rate(RR)by using the temperature of exhaled *** hypothesized that the temperature sensor is at least equivalent in accuracy to capnography in monitoring *** To determine if the temperature sensor is equivalent to capnography in monitoring procedural *** In this prospective study,participants were individually fitted with a face mask monitored by both LRMD and *** following data were collected:gender,age,body mass index,type of procedure,and doses of *** each patient,we report the mean RR for each device as well as the mean *** analyses were performed using SAS,and a P0.05 was considered statistically *** Twelve consecutive patients undergoing endoscopic procedures at our institution were *** patients were excluded due to incomplete data,inadequate data,patient cooperation,and capnography ***,we found that LRMD RR highly correlated to capnography RR(P0.001);the average capnography RR increases by 0.66 breaths for every one additional breath measured by the *** addition,apnea rates were 7.4%for the capnography and 6.4%for the LRMD(95%confidence interval:0.92-1.10).CONCLUSION The LRMD correlated with the gold standard capnography with respect to respiratory rate detection and apnea *** LRMD could be used as an alternative to capnography for measuring respiration in endoscopy.