Mid-Term Review of ADEPT Metal-On-Metal Hip Prosthesis. Functional, Radiological and Metal Ion Analysis

Mid-Term Review of ADEPT Metal-On-Metal Hip Prosthesis. Functional, Radiological and Metal Ion Analysis

作     者:James G. A. Plant Gareth H. Prosser Ben J. Burston Stephen J. Edmondston Piers J. Yates 

作者机构:Department of Orthopaedics Fremantle Hospital Fremantle Australia 

出 版 物:《Open Journal of Orthopedics》 (矫形学期刊(英文))

年 卷 期:2014年第4卷第2期

页      面:38-43页

学科分类:1002[医学-临床医学] 100214[医学-肿瘤学] 10[医学] 

主  题:Metal-On-Metal Arthroplasty Modular Resurfacing ADEPT Chromium Ions Cobalt Ions Functional Review 

摘      要:Background: There is much interest regarding metal-on-metal implants in medical and general media. Much of this has been regarding failure of specific implant systems and metal ion toxicity. We present our early mid-term experience of the ADEPT metal-on-metal system which has both modular and non-modular hip options. Methods: Functional assessment, blood metal ion quantification, and radiographic analysis were performed for the modular and non-modular ADEPT variants. Fifty implants were implanted with a mean follow up time of 28 months. Unpaired t-tests were used to compare modular and resurfacing groups, standardised hip scores were used to compare function to conventional total hip arthroplasty. Results: Metal ion levels were significantly higher in modular prostheses compared to resurfacing implants, but not at “harmful levels (as determined by a previous metal-on-metal implant recall). Functional outcomes were excellent and revision rates were lower than expected. Conclusions: At our institution we have good outcomes with the ADEPT hip prosthesis. Though patient selection and implant position are crucial, poor performance of metal-on-metal hip replacements is implant specific.

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分