Feasibility of Localizing in vivo Micro-devices with Electromagnetic Methods
Feasibility of Localizing in vivo Micro-devices with Electromagnetic Methods作者机构:School of Electronic Information and Electrical Engineering Shanghai Jiaotong University
出 版 物:《Journal of Shanghai Jiaotong university(Science)》 (上海交通大学学报(英文版))
年 卷 期:2008年第13卷第5期
页 面:559-561页
核心收录:
学科分类:080802[工学-电力系统及其自动化] 0808[工学-电气工程] 08[工学] 080401[工学-精密仪器及机械] 0804[工学-仪器科学与技术] 080402[工学-测试计量技术及仪器]
基 金:the Specialized Research Fundfor the Doctoral Program of Higher Education(No. 20040248033) the National High Technology Research and Development Program (863) of China(No. 2006AA04Z368)
主 题:medical micro-devices position measurement electromagnetic field
摘 要:In order to measure the location of medical micro-devices in the gastrointestinal tract, electro-magnetic methods were developed considering the magnetic permeability of human bodies. Two varieties of electromagnetic localization methods have been implemented: one uses direct current(DC) to generate the elec-tromagnetic field, and the other uses alternating current (AC). The two varieties of the localization principle were analyzed and relevant experiments were made. The experiments show that the alternating electromagnetic method acquires higher accuracy, higher stability and wider measurement range.