An experimental inter-comparison of radiotherapic dosimetric delivery to assess X-Ray linear accelerator output characteristics

基于放射疗法剂量给药对比的X射线直线加速器输出特性评估(英文)

作     者:K Masood T Zafar J Zafar H Zafar 

作者机构:Institute of Nuclear Medicine & Oncology (INMOL) Department of EconomicsGovernment College University Faculty of EngineeringDepartment of Electrical EngineeringGovernment College University Tissue Optics & Microcirculation Imaging Research GroupNational University of Ireland 

出 版 物:《Journal of Measurement Science and Instrumentation》 (测试科学与仪器(英文版))

年 卷 期:2014年第5卷第3期

页      面:83-86页

学科分类:0831[工学-生物医学工程(可授工学、理学、医学学位)] 1004[医学-公共卫生与预防医学(可授医学、理学学位)] 08[工学] 10[医学] 

主  题:X-ray linear accelerator dosimetric analysis absorbed chamber phantom 

摘      要:The dosimetric inter-comparison studies carry pertinent significance to ensure uniform radiation dose delivery for clinical trials. This paper investigates the comparative performance analysis of an X-ray high power linear accelerator per- formed by the International Dosimetry Survey Mission and Institute of Nuclear Medicine & Oncology (INMOL), Lahore. The measurements were made using cylindrical ionization chambers based on the International Atomic Energy Agency (IAEA) TRS-398 protocol for absorbed dose-to-water dosimetric standards, and the percentage deviation was found to be between 0.5 % - 1%. The dosimetric analysis concerning linear X-ray accelerator output performed by INMOL, Lahore was found to be in good agreement with the results of IAEA Dosimtery Survey Mission.