咨询与建议

看过本文的还看了

相关文献

该作者的其他文献

文献详情 >Determination of Epinephrine... 收藏

Determination of Epinephrine Using Poly(Neutral Red)Modified Carbon Fibre Microelectrodes by Anodic Stripping Voltammetry

Determination of Epinephrine Using Poly(Neutral Red) Modified Carbon Fibre Microelectrodes by Anodic Stripping Voltammetry

作     者:Tang Xiaorong, Sun Yuanxi, Yao Bing, He Fei, Zhou Xingyao Department of Chemistry, Wuhan University, Wuhan 430072, China 

出 版 物:《Wuhan University Journal of Natural Sciences》 (武汉大学学报(自然科学英文版))

年 卷 期:1997年第2卷第4期

页      面:97-101页

学科分类:071011[理学-生物物理学] 0710[理学-生物学] 0401[教育学-教育学] 04[教育学] 07[理学] 

主  题:carbon fibre microelectrode poly(neutral red) epinephrine anodic stripping voltammetry 

摘      要:Poly(neutral red) film modified carbon fibre microelectrodes offer substantial improvement in voltammetric sensitivity and selectivity towards epinephrine (EP). The poly(neutral red) film was electropolymerized by cycling the potential between -0.8 V and +0.8 V. The anodic stripping voltammetric response for EP was found to be dependent on accumulation time and potential. By using a poly(neutral red) film modified carbon fibre microelectrode with a 1 min preconcentration at -1.2 V in biological phosphate buffer solution (pH=7.4), a good linear relationship between the anodic stripping peak currents and EP concentrations was obtained in the range of 2.0×10 -7  ̄ 2.0×10 -5 mol/L. The detection limit was up to 9.0 ×10 -8 mol/L. Moreover, 400 times higher concentrations of vitamin C did not interfere with the measurement of EP. This method was used for determining EP concentrations in epinephrine hydrochloride injection solution with satisfactory results.

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分