Methods for obtaining characteristic γ-ray net peak count from interlaced overlap peak in HPGe γ-ray spectrometer system
Methods for obtaining characteristic γ-ray net peak count from interlaced overlap peak in HPGe γ-ray spectrometer system作者机构:School of Electrical and Mechanical EngineeringPingdingshan University
出 版 物:《Nuclear Science and Techniques》 (核技术(英文))
年 卷 期:2019年第30卷第1期
页 面:73-78页
核心收录:
学科分类:082704[工学-辐射防护及环境保护] 08[工学] 0827[工学-核科学与技术]
基 金:supported by the National Natural Science Foundation of China(Nos.11575090,11605099) the Young Key Teachers Training Program of He’nan Higher Education in China(No.2015GGJS-258)
主 题:Peak count Interlaced overlap peak Highpurity germanium (HPGe)γ-ray spectrometer system
摘 要:For a characteristic c-ray with interlaced overlap peak, and the case where its reliable and credible net count cannot be obtained using the current high-purity germanium(HPGe) multichannel γ-ray spectrum software, two new methods are proposed herein to obtain the γ-ray net peak count from the interlaced overlap peak in the HPGe cray spectrometer system, of which one is the symmetric conversion method based on Gaussian distribution and the other is where the energy average value of two close γ-rays is regarded as the γ-ray energy. The experimental results indicate that the two methods mentioned above are reliable and credible. This study is significant for the development of better γ-ray spectrum processing software for measuring complex γ-ray spectra concerning the nuclear reaction cross section, neutron activation analysis, and analysis of transuranium elements, using an HPGe detector.