Modeling and Simulation of Human Cornea for the Measurement of Intraocular Pressure (IOP) through Eyelid

作     者:Mahesh B.Mawale Abhaykumar Kuthe Padma G.Pawane Sandeep W.Dahake 

作者机构:Kavikulguru Institute of Technology and ScienceRamtekNagpurMaharashtra 441106India Visvesvaraya National Institute of TechnologyNagpurMaharashtra 440001India ESIS HospitalNagpurMaharashtra 440001India 

出 版 物:《Computer Modeling in Engineering & Sciences》 (工程与科学中的计算机建模(英文))

年 卷 期:2018年第114卷第2期

页      面:173-189页

核心收录:

学科分类:1002[医学-临床医学] 100210[医学-外科学(含:普外、骨外、泌尿外、胸心外、神外、整形、烧伤、野战外)] 10[医学] 

主  题:Glaucoma cornea sclera eyelid intraocular pressure tonometry 

摘      要:This article describes a study based on the use of an instrumented form of applanation and indentation tonometry through *** methods and devices which are available for the measurement of intraocular pressure(IOP)have their own *** cause discomfort to the patients during IOP *** these methods,used to measure IOP over cornea,need *** the present work,an attempt is made to develop a device for the measurement of IOP through eyelid which eliminates the need of anesthesia and enhance the patient’s comfort and convenience of the *** represent the results of proposed technique,the human eye modeled with cornea,sclera and also with the eyelid,is ***,simulation results of stress values for the range of IOP i.e.5-40 mm of Hg are presented and discussed comparatively with eyelid and without eyelid ***‘on cornea’methods.