Ocuton S眼压计自测法:青光眼患者的眼压测量准确性评价

Self-tonometry with the Ocuton S:Evaluation of accuracy in glaucoma patients

作     者:Sacu S. Vass C. Schemper M. Rainer G. 杨建刚 

作者机构:Medical University Vienna Department of Ophthalmology Waehringer Guertel 18-20 A-1090 Vienna Austria Dr. 

出 版 物:《世界核心医学期刊文摘(眼科学分册)》 (Digest of the World Core Medical Journals:Ophthalmology)

年 卷 期:2005年第1卷第2期

页      面:6-6页

学科分类:1002[医学-临床医学] 100212[医学-眼科学] 10[医学] 

主  题:眼压测量 Ocuton S 角膜厚度 角膜地形图 重复测量 仪器检查 

摘      要:Background: To compare two methods of measuring intraocular pressure (IOP) and to evaluate whether repeated measurements taken with the Ocuton S applanation s elf-tonometer can improve reliability. Methods: Ocuton S and Goldmann applanati on tonometry (GAT), and corneal thickness measurements taken with the Orbscan to pography system, were successfully performed in 64 of 68 glaucoma patients. Resu lts: The median IOPs were 15.5mmHg using GAT, and 16 mmHg using the first self- taken Ocuton S measurement (n=64). The differences between the median of the GAT measurements and the first Ocuton S measurement, and the medians of the three a nd six separate Ocuton S measurements were within 3 mmHg in 52%, 59%and 67%of cases, respectively. The mean corneal thickness of all evaluated eyes was 545.3 μm. There was no effect of corneal thickness on the accuracy of either of the two devices (P 0.05). Conclusion: Repeated measurements can improve the reliab ility of the Ocuton S. However, ever with repeated measurements only every secon d patient succeeds in obtaining reliable measurements.