咨询与建议

看过本文的还看了

相关文献

该作者的其他文献

文献详情 >Enantioseparation of Palonoset... 收藏

Enantioseparation of Palonosetron Hydrochloride and Its Related Enantiomeric Impurities by Computer Simulation and Validation

盐酸帕洛诺司琼对映体及其相关的对映体杂质的计算机模拟和验证

作     者:M.Vishnu Murthy C.Krishnaiah Kodithyala Jyothirmayi Katkam Srinivas K.Mukkanti Ramesh Kumar Gautam Samanta 

作者机构:Analytical Research and DevelopmentIntegrated Product Development OperationsDr.Reddy’s Laboratories Ltd.BachupallyQutubullapurAndhra PradeshIndia Department of ChemistryJawaharlal Nehru Technological UniversityHyderabadIndia 

出 版 物:《American Journal of Analytical Chemistry》 (美国分析化学(英文))

年 卷 期:2011年第2卷第4期

页      面:437-446页

学科分类:07[理学] 0703[理学-化学] 

主  题:High Performance Liquid Chromatography Palonosetron Hydrochloride Chiral Impurities Drylab^(█) Software Method Development Method Validation 

摘      要:A rapid, simple and single stereo selective high-performance liquid chromatographic (HPLC) method was developed and validated for enantiomers of palonosetron hydrochloride (PALO) and its process related chiral impurities. A computer simulating software was used for the development of chiral method. The developed method was able to separate not only the enantiomers of palonosetron hydrochloride but also its process related chiral impurities within 12 min. The chromatographic separation was carried out by normal phase chromatography using a 3 μm column of cellulose based chiral stationary phase (Chiralcel-OD 250mm × 4.6mm) with a mobile phase comprised of n-hexane: ethanol: methanol: heptafluoro butyric acid: diethyl amine (70:15:15:0.05:0.1, v/v) at a flow rate of 1.0 mL/min. The effects of additive concentration as well as nature of polar organic modifier, flow rate, and temperature on enantioselectivity were investigated. The limit of detection (LOD) and limit of quantification (LOQ) of the palonosetron isomers and its related chiral impurities were found to be in the range 0.06-0.10 μg/mL and 0.14 - 0.24 μg/mL respectively. The method showed excellent linearity (R2 0.998) over a range of 0.14 to 1.125 μg/mL. The percentage recovery of the isomers in bulk drug samples ranged from 87.0 to 116.0.

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分