Verification and Dosimetric Impact of Acuros XB Algorithm for Stereotactic Body Radiation Therapy (SBRT) and RapidArc Planning for Non-Small-Cell Lung Cancer (NSCLC) Patients
Verification and Dosimetric Impact of Acuros XB Algorithm for Stereotactic Body Radiation Therapy (SBRT) and RapidArc Planning for Non-Small-Cell Lung Cancer (NSCLC) Patients作者机构:Department of Medical Physics ProCure Proton Therapy Center Oklahoma City USA Department of Radiation Oncology Arizona Center for Cancer Care Peoria USA
出 版 物:《International Journal of Medical Physics, Clinical Engineering and Radiation Oncology》 (医学物理学、临床工程、放射肿瘤学(英文))
年 卷 期:2013年第2卷第1期
页 面:6-14页
学科分类:1002[医学-临床医学] 100214[医学-肿瘤学] 10[医学]
主 题:Acuros XB AAA Heterogeneity Correction SBRT RapidArc Lung Cancer
摘 要:Purpose: The experimental verification of the Acuros XB (AXB) algorithm was conducted in a heterogeneous rectangular slab phantom, and compared to the Anisotropic Analytical Algorithm (AAA). The dosimetric impact of the AXB for stereotactic body radiation therapy (SBRT) and RapidArc planning for 16 non-small-cell lung cancer (NSCLC) patients was assessed due to the dose recalculation from the AAA to the AXB. Methods: The calculated central axis percentage depth doses (PDD) in a heterogeneous slab phantom for an open field size of 3 ×3 cm2 were compared against the PDD measured by an ionization chamber. For 16 NSCLC patients, the dose-volume parameters from the treatment plans calculated by the AXB and the AAA were compared using identical jaw settings, leaf positions, and monitor units (MUs). Results: The results from the heterogeneous slab phantom study showed that the AXB was more accurate than the AAA;however, the dose underestimation by the AXB (up to ?3.9%) and AAA (up to ?13.5%) was observed. For a planning target volume (PTV) in the NSCLC patients, in comparison to the AAA, the AXB predicted lower mean and minimum doses by average 0.3% and 4.3% respectively, but a higher maximum dose by average 2.3%. The averaged maximum doses to the heart and spinal cord predicted by the AXB were lower by 1.3% and 2.6% respectively;whereas the doses to the lungs predicted by the AXB were higher by up to 0.5% compared to the AAA. The percentage of ipsilateral lung volume receiving at least 20 and 5 Gy (V20 and V5 respectively) were higher in the AXB plans than in the AAA plans by average 1.1% and 2.8% respectively. The AXB plans produced higher target heterogeneity by average 4.5% and lower plan conformity by average 5.8% compared to the AAA plans. Using the AXB, the PTV coverage (95% of the PTV covered by the 100% of the prescribed dose) was reduced by average 8.2% than using the AAA. The AXB plans required about 2.3% increment in the number of MUs in order to achieve the sa