Effects of oral contrast on dose in abdominopelvic computed tomography with pure iterative reconstruction
Effects of oral contrast on dose in abdominopelvic computed tomography with pure iterative reconstruction作者机构:Department of Radiology Cork University Hospital
出 版 物:《World Journal of Radiology》 (世界放射学杂志(英文版)(电子版))
年 卷 期:2016年第8卷第9期
页 面:809-815页
学科分类:0831[工学-生物医学工程(可授工学、理学、医学学位)] 100207[医学-影像医学与核医学] 1006[医学-中西医结合] 1002[医学-临床医学] 1001[医学-基础医学(可授医学、理学学位)] 08[工学] 1010[医学-医学技术(可授医学、理学学位)] 100201[医学-内科学(含:心血管病、血液病、呼吸系病、消化系病、内分泌与代谢病、肾病、风湿病、传染病)] 100106[医学-放射医学] 1009[医学-特种医学] 10[医学] 100602[医学-中西医结合临床]
主 题:Radiation dose Low dose computed tomography Abdominal imaging Oral contrast Computed tomography
摘 要:AIM To assess the effect of neutral(NC) and positive(PC) oral contrast use on patient dose in low-dose abdominal computed tomography(CT).METHODS Low-dose clinically indicated CTs were performed on 79 Crohn s patients(35 = PC, 1 L 2% gastrografin; 44 = NC, 1.5 L polyethylene glycol). Scanner settings for both acquisitions were identical apart from 25 s difference in intravenous contrast timing. Body mass index(BMI), scan-ranges, dose-length product and size-specific dose estimated were recorded. Data was reconstructed with pure model-based iterative reconstruction. Image quality was objectively and subjectively analysed. Data analysis was performed with Statistical Package for Social *** Higher doses were seen in neutral contrast CTs(107.60 ± 78.7 m ***, 2.47 ± 1.21 m Gy vs 85.65 ± 58.2 m ***, 2.18 ± 0.96 m Gy). The differencehad both NC and PC investigations. Image-quality assessment yielded 6952 datapoints. NC image quality was significantly superior(P 0.001)(objective noise, objective signal to noise ratio, subjective spatial resolution, subjective contrast resolution, diagnostic acceptability) at all levels. NC bowel distension was significantly(P 0.001) *** The use of polyethylene glycol as a neutral OC agent leads to higher radiation doses than standard positive contrast studies, in low dose abdominal CT imaging. This is possibly related to the osmotic effect of the agent resulting in larger intraluminal fluid volumes and resultant increased overall beam attenuation.