A FLUORESCENCE QUENCHING IMNUNOASSAY METHOD FOR DETERMINATION OF TRACE ALBUMIN USING UNLABELED EUROPIUM CHELATE
A FLUORESCENCE QUENCHING IMNUNOASSAY METHOD FOR DETERMINATION OF TRACE ALBUMIN USING UNLABELED EUROPIUM CHELATE出 版 物:《Chinese Chemical Letters》 (中国化学快报(英文版))
年 卷 期:1992年第3卷第12期
页 面:1011-1014页
核心收录:
基 金:This work was supported by National Natural Science Foundation of China
主 题:RSA Eu A FLUORESCENCE QUENCHING IMNUNOASSAY METHOD FOR DETERMINATION OF TRACE ALBUMIN USING UNLABELED EUROPIUM CHELATE
摘 要:A fluoroimmunoassay method using unlabeled europium chelate is *** principle is similar to that of fluoroimmunoassay method using lanthanide chelate as *** procedure is simple because labeling process is *** detection limit is about 10^(10) mol/L *** relative standard deviation of immunoassay is less than 10%.The recoveries of human serum albumin and estradiol protein conjugate are 96-105% and 111% respectively.