咨询与建议

看过本文的还看了

相关文献

该作者的其他文献

文献详情 >Metallosis with spinal implant... 收藏

Metallosis with spinal implant loosening after spinal instrumentation:A case report

作     者:Yiu Hin Kwan Hong Lee Terry Teo Shree Kumar Dinesh Wee Lim Loo 

作者机构:Department of Orthopaedic SurgeryChangi General HospitalSingapore 529889Singapore 

出 版 物:《世界骨科杂志:英文版》 (World Journal of Orthopedics)

年 卷 期:2023年第14卷第8期

页      面:651-661页

学科分类:1002[医学-临床医学] 100210[医学-外科学(含:普外、骨外、泌尿外、胸心外、神外、整形、烧伤、野战外)] 10[医学] 

主  题:Metallosis Spine Instrumentation Implant loosening Corrosion Case report 

摘      要:BACKGROUND Spinal metallosis is a rare complication following spinal instrumentation whereby an inflammatory response to the metal implants results in the development of granulomatous *** SUMMARY We describe the case of a 78-year-old woman who had recurrence of back pain 5 years after lumbar spine posterior decompression and instrumented *** spine radiographs showed hardware loosening and magnetic resonance imaging showed adjacent segment *** surgery revealed evidence of metallosis *** Spinal metallosis can present several years after *** and computed tomography may demonstrate hardware loosening secondary to *** metal concentrations associated with spinal metallosis have yet to be ***,metallosis is still an intraoperative and histopathological *** presence of metallosis after spinal instrumentation likely indicates a more complex underlying problem:Pseudarthrosis,failure to address sagittal balance,infection,and cross-threading of set ***,identifying metallosis is important,but initiating treatment promptly for symptomatic implant loosening is of greater paramount.

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分