Image-Based Ultrasound Speed Estimation: Phantom and Human Liver Studies
Image-Based Ultrasound Speed Estimation: Phantom and Human Liver Studies作者机构:Department of Radiology and Medical Imaging Stritch School of Medicine Loyola University Medical Center Chicago Illinois USA
出 版 物:《Open Journal of Radiology》 (放射学期刊(英文))
年 卷 期:2023年第13卷第2期
页 面:101-112页
学科分类:08[工学] 0812[工学-计算机科学与技术(可授工学、理学学位)]
主 题:Ultrasound Image Normalized Autocorrelation Function (ACF) Speed of Sound (SoS)
摘 要:Purpose: A novel image-based method for speed of sound (SoS) estimation is proposed and experimentally validated on a tissue-mimicking ultrasound phantom and normal human liver in vivo using linear and curved array transducers. Methods: When the beamforming SoS settings are adjusted to match the real tissue’s SoS, the ultrasound image at regions of interest will be in focus and the image quality will be optimal. Based on this principle, both a tissue-mimicking ultrasound phantom and normal human liver in vivo were used in this study. Ultrasound image was acquired using different SoS settings in beamforming channels ranging from 1420 m/sec to 1600 m/sec. Two regions of interest (ROIs) were selected. One was in a fully developed speckle region, while the other contained specular reflectors. We evaluated the image quality of these two ROIs in images acquired at different SoS settings in beamforming channels by using the normalized autocorrelation function (ACF) of the image data. The values of the normalized ACF at a specific lag as a function of the SoS setting were computed. Subsequently, the soft tissue’s SoS was determined from the SoS setting at the minimum value of the normalized ACF. Results: The value of the ACF as a function of the SoS setting can be computed for phantom and human liver images. SoS in soft tissue can be determined from the SoS setting at the minimum value of the normalized ACF. The estimation results show that the SoS of the tissue-mimicking phantom is 1460 m/sec, which is consistent with the phantom manufacturer’s specification, and the SoS of the normal human liver is 1540 m/sec, which is within the range of the SoS in a healthy human liver in vivo. Conclusion: Soft tissue’s SoS can be determined by analyzing the normalized ACF of ultrasound images. The method is based on searching for a minimum of the normalized ACF of ultrasound image data with a specific lag among different SoS settings in beamforming channels.