Actual Use of Rigid Hysteroscope Intrauterine BIGATTI Shaver (IBS<sup>®</sup>) with a Morcellator System

Actual Use of Rigid Hysteroscope Intrauterine BIGATTI Shaver (IBS<sup>®</sup>) with a Morcellator System

作     者:Toshiyuki Kakinuma Kaoru Kakinuma Ayano Shimizu Takahum Okusa Eri Suizu Masataka Kagimoto Masaki Abe Koyomi Saito Michitaka Ohwada Nobuhiro Takeshima Kaoru Yanagida Toshiyuki Kakinuma;Kaoru Kakinuma;Ayano Shimizu;Takahum Okusa;Eri Suizu;Masataka Kagimoto;Masaki Abe;Koyomi Saito;Michitaka Ohwada;Nobuhiro Takeshima;Kaoru Yanagida

作者机构:Department of Obstetrics and Gynecology International University of Health and Welfare Hospital Iguchi Nasushiobara-City Tochigi Japan 

出 版 物:《Open Journal of Obstetrics and Gynecology》 (妇产科期刊(英文))

年 卷 期:2021年第11卷第11期

页      面:1494-1499页

学科分类:1002[医学-临床医学] 100214[医学-肿瘤学] 10[医学] 

主  题:Hysteroscopic Surgery Morcellator Uterine Submucosal Fibroids Endometrial Polyps 

摘      要:Hysteroscopic resection of endometrial lesions has recently gained popularities. We here report the usefulness and safety of hysteroscopic surgery using Intrauterine BIGATTI Shaver (IBS®). We performed this surgery for 6 patients: 4 endometrial polyps and 2 intrauterine myoma. The average operation time was 25.7±13.2 minutes. Only a small amount of bleeding was noted in all cases. There were no surgical complications. Two out of four patients who desired childbearing became pregnant. Hysteroscopic surgery using IBS® can safely and reliably remove intrauterine protruding lesions under aspiration without using a heat source. Less endometrial damage may help preserve fertility.